This page assesses the risks specific to the v3 cradle. It is a passive, non-powered, 3D-printed part built on the OpenPoleMount mounting block.

General-purpose hardware

The v3 cradle is general-purpose mounting hardware — not a medical device. No electrical, chemical, or software function; no patient contact; no role in delivering or controlling therapy. The only hazard category is mechanical. See the Safety & Disclaimer.

Shared clamp risks

The v3 cradle attaches with the same block-and-thumbscrew clamp as every OpenPoleMount design. Those load-bearing hazards (slip, overload, thumbscrew over-tightening, weak/mis-oriented print, material creep) and their mitigations are covered once in the Mounting Block Risk Analysis — read it alongside this page.

Load rating

Recommended working load: up to ~1.4 kg (3 lb), as a conservative starting point at the recommended print settings. Verify safe support of your own load first.

v3-specific hazards

# Hazard Cause Effect Mitigation
1 Object does not seat fully Object larger than the form-fitting pocket Poor seating; object sits high or proud v3's pocket is sized to a compact envelope (~124 × 99 × 65 mm). Check your object fits before relying on it; use v2 for a more open pocket, or resize the model.
2 Object rocks in the pocket Object smaller than the pocket Movement under load The form-fitting v3 pocket minimises this vs. an open holder, but confirm a snug fit; add a shim or reprint scaled if loose.
3 Front face obstructed Object oriented wrong way in the pocket Controls/screen not reachable Seat the object with its front face outward — the v3 pocket is shaped to keep the front accessible.
4 Cabling strain Cable/tubing routed under tension Pull on the mount or the object Route cabling with slack so it exits without tension (see Assembly).

Evidence

The load-bearing clamp shared by v3 was validated in destruction testing: a deliberately under-spec print never failed under normal load, the failure mode was safe and visible, and the broken mount retained residual grip. The v3 body reuses that same clamp geometry; its form-fitting pocket has not been separately tested to failure.

Points of failure

Point Consequence Detectability Net risk
Block/thumbscrew clamp See block analysis Visible; residual grip Low with correct install
Pocket fit Object rocks or sits proud if mismatched Felt on test-fit and load test Low with correct-size object
Print integrity Weak part if settings/orientation wrong Inspection + hand flex-test Low with adherence to print guide

Informational only; not engineering certification or regulatory approval. See the Safety & Disclaimer.