Safety Note — Read First

This is a load-bearing mount, rated for equipment up to approximately 1.4 kg (3 lbs). Print quality directly affects safety. The settings below are minimums — you are responsible for verifying mechanical integrity before use. See Safety & Disclaimer.

STL Files Available Now

The finalized OpenPoleMount_IV-Pole_v3.stl is published and print-ready — download it directly or from GitHub Releases.

v2 also remains fully released and print-ready.

V3 vs V2 — What's Different

IV Pole Mount v3 uses a smaller, more form-fitting pocket (~125 × 100 × 66 mm printed), redesigned from feedback on v2's more open pocket. The pole channel and thumbscrew interface are unchanged. Print settings below carry over the same conservative minimums validated for v1/v2 — the v3 pocket itself has not yet been separately destruction-tested. Treat these as a starting point, not a substitute for your own inspection before use.

Required Files

File Purpose Quantity
OpenPoleMount_IV-Pole_v3.stl IV pole cradle (v3 shape) 1
thumbscrew_v14_flat_Thandle.stl Thumbscrew for pole attachment (same part as v1/v2) 1

Testing Status — Pending

The destruction testing on file was performed on the v1/v2 cradle body shape. v3 has not yet been separately tested to failure. The settings below are carried over from that testing as a conservative starting point because the pole channel, thumbscrew boss, and wall construction method are shared across all three versions — but this has not been independently confirmed for the v3 pocket geometry specifically. (Pending George's confirmation/testing before this note can be upgraded to a tested claim.)

Material

PLA is the tested and validated material for this design family. Settings on this page are based on PLA testing of v1/v2. Other materials (PETG, ASA, ABS) have not yet been tested — if you print in another material, please share your results by opening an issue.

Material Status
PLA ✅ Tested (v1/v2 geometry) — use the settings below
PETG / ASA / ABS ⚠️ Not yet tested — proceed at your own risk
TPU / Flexible ❌ Not suitable — cannot support structural load
Setting Cradle Thumbscrew
Layer height 0.2 mm 0.2 mm
Wall count 6 walls (≈2.4 mm at 0.4 mm nozzle) 6 walls
Top/bottom thickness 2 mm minimum 2 mm minimum
Infill 35% minimum — 3D honeycomb (preferred) or other non-concentric pattern 35% minimum
Infill pattern 3D honeycomb (preferred), gyroid, or cubic — not lines, grid, or concentric same
Supports None needed Optional — single layer keeps part in place; brim works too
Print speed 40–60 mm/s 40–60 mm/s

Wall and Infill Minimums Are Safety-Critical

Testing on the v1/v2 shape has shown that the mount can crack if over-tightened when walls or infill are below these minimums. The 6-wall + 2 mm top/bottom + 35% non-concentric infill combination is the recommended minimum until v3-specific testing is complete.

Do not use concentric, lines, or grid infill patterns — they do not provide the same resistance to the stress concentration from the thumbscrew. 3D honeycomb is the preferred pattern.

Part Orientation

Cradle: Print with the pole channel vertical (mounting hole facing up). This orients layer lines perpendicular to the primary load direction, maximizing pull-out and shear strength.

Thumbscrew: Print flat with the T-handle face down. No supports needed.

Support Settings

Neither part requires supports. Supports are optional on both:

  • Cradle: supports are not required. If you use them, place under the equipment cradle pocket. Tree or Normal both work; 45° threshold; 2–3 interface layers for clean removal.
  • Thumbscrew: supports are not required. If enabled, the slicer typically places a single layer under the part — this helps hold it in place during printing. A brim works equally well for this purpose.

Post-Processing

After printing:

  1. Remove all supports carefully — use flush cutters, not a knife
  2. Clean the thumbscrew threads with a brush if needed
  3. Test-fit thumbscrew in the threaded hole — should turn smoothly by hand
  4. Verify the pole channel diameter matches your IV pole
  5. Check device fits snugly in the cradle pocket — should not rattle

Verification Before Use

Run through this checklist before mounting any equipment:

  • [ ] No visible layer delamination (check by pressing firmly — should not crack or flex excessively)
  • [ ] No significant warping of the base or pole channel
  • [ ] Thumbscrew threads cleanly with no binding or stringing
  • [ ] Dry test: mount on pole without equipment, apply firm downward pressure — cradle must not slide
  • [ ] Device seats fully without rocking or side-to-side movement — pending George's own physical fit-test report on the finished v3 print; the photos on the overview page show a successful hand-fit but have not been logged as a formal verification pass

If any check fails, do not use the part — reprint with corrected settings.

Over-Tightening Risk

Once mounted, tighten the thumbscrew hand-tight plus one quarter-turn only. Over-tightening is the primary cause of cradle failure even in correctly printed parts. See the Assembly Guide for tightening instructions.